HPLC

LC-2008 UV Detector

Specifications:

Parameter Specification
Wavelength Range 190-600nm
Bandpass 8nm
Wavelength Accuracy Better than ±2nm
Wavelength Reproducibility Better than 0.2nm
Baseline Noise ≤ ±5 X 10-5 AU (MeOH, 1mL/min, 254nm, 20ºC)
Baseline Drift ≤ ±5 X 10-4 AU/hr (MeOH, 1mL/min, 254nm, 20ºC)
Minimum Inspection Data 3 X 10 -8 g/ml (naphthalene in MeOH )
Minimal Concentration ≤ 2 X 10 -8 g/ml (naphthalene in MeOH )
Linearity Range Better than 104
ABS Range (AU/FS) 0.001, 0.005, 0.01, 0.05, 0.1, 0.5, 1, 2
Filter Time Constant 0.1, 0.2, 0.5, 1, 2
Flow Cell Volume 8 microlitre (8μl)
Fuse 2A
Analog Output 10mV
Recorder Output 10mV
Data Processor Output 1V
Remote Startup Control Connected with manual sampling
S-485 Data Communication Input Connected with pump
RS-232 Data Communication Input Connected with software
Dimensions 450mm L X 300mm W X 160mm H
Net Weight 12Kg (26.46lbs)